Free SCFHS Biomedical Engineering & Medical Devices mock test

25 original exam-style questions from our Biomedical Engineering & Medical Devices bank, timed at the exam’s own pace, scored against the 60% SCFHS pass mark. They span 10 of the 12 blueprint domains, so you will see exactly where you are weakest - free, no sign-up.

Pass mark SCFHS 60%

Free diagnosticNo sign-up · no email · results instantly

Sit 25 original exam-style questions at real exam pace

These are 25 original questions taken from our Biomedical Engineering & Medical Devices bank, spread across 10 of the 12 domains of the published blueprint. You get 72 seconds per question, which is exactly the pace of the real exam (30 minutes for 25). At the end you will see which domains you are weakest in, and the full worked reasoning for every question.

25
questions
30:00
time limit
60%
SCFHS pass mark

A short diagnostic cannot predict whether you will pass - only the real exam does that. It will show you, honestly, where you currently stand.

Quick answer

The Biomedical Engineering & Medical Devices exam at a glance

The Biomedical Engineering & Medical Devices licensing exam is a computer-based, multiple-choice test, and each GCC regulator sets its own. SCFHS publishes exact figures, listed below. This free mock gives you 25 original exam-style questions at 72 seconds each, scored against the 60% SCFHS pass mark, then shows your weakest domains.

Published exam figures

  • SCFHS

    Saudi Classification Examination - Biomedical Engineering (Medical Devices) Specialist

    Saudi Arabia150 MCQs180 minutes

    60%

    pass mark

Attempts in the UAE
3 in total across DHA, DOH, MOHAP and SHA
This free mock
25 questions · 30 minutes · 10 of 12 domains
Full bank
450+ questions · 3 to 6 papers · AED 269
Regulators
SCFHS

How this free Biomedical Engineering & Medical Devices mock works

25
original questions
30 min
time limit
72s
per question
10 of 12
blueprint domains
  • You get 72 seconds per question, the pace of a 150-question paper in 3 hours. The timer runs in your browser and ends the sitting when it reaches zero.
  • Your score is set against the 60% SCFHS pass mark, the published figure for this exam. No other number is invented: there is no percentile and no pass probability.
  • At the end you see your score, your pace per question, every domain ranked weakest first, and the worked answer and source for all 25 questions.
  • You can mark each answer “sure” or “not sure”. Wrong answers you were sure about are shown first, because they are the ones you would not stop to check in the real exam.
  • Your last result is saved in your own browser only, so you can come back and compare. Nothing is sent to us.

What the real Biomedical Engineering & Medical Devices licensing exam looks like

The Biomedical Engineering & Medical Devices exam is a computer-based multiple-choice paper, published by one regulator: SCFHS (pass mark 60%). Every regulator's format, pass mark and fee is in the table below; confirm yours before booking.

Pass marks at a glance: exam codes and pass marks by regulator for every written exam in one sourced table.

Before you can book the exam at all you need Dataflow primary source verification of your degree, licence and experience, which typically runs several weeks. Candidates often underestimate this and end up sitting the exam sooner after study than they planned. Check where you stand with our free eligibility checker or confirm whether your title even requires an exam using the Prometric exam checker.

You get three attempts in total across all UAE authorities, not three each, under the Unified Professional Qualification Requirements. At roughly AED 880 per attempt, plus the licensing delay each failure adds to your visa and start date, the cost of going in underprepared is not really the exam fee.

Biomedical Engineering & Medical Devices licensing exam in every GCC country
RegulatorExamFormatPass markFee per attempt
SCFHS
Kingdom of Saudi Arabia
Practise →
Saudi Classification Examination - Biomedical Engineering (Medical Devices) Specialist
150 MCQs in 180 minutes60%Not published
No exam published for this title
Not in the DHA CBT list; DHA's guideline assesses licensed titles that are not listed by oral
Not publishedNot publishedNot published
DOH
Abu Dhabi, Al Ain and Al Dhafra
Practise →
No exam published for this title
DOH publishes no exam for this title that we could find; check with DOH
Not publishedNot publishedNot published
MOHAP
Northern Emirates - Ajman, Umm Al Quwain, Ras Al Khaimah and Fujairah
Practise →
No exam published for this title
MOHAP publishes no exam for this title that we could find; check with MOHAP
Not publishedNot publishedNot published
SHA
Emirate of Sharjah
Practise →
No exam published for this title
SHA publishes no exam for this title that we could find; check with SHA
Not publishedNot publishedNot published
QCHP
State of Qatar
Practise →
No exam published for this title
No DHP Qatar exam blueprint published for this title; check with DHP
Not publishedNot publishedNot published
NHRA
Kingdom of Bahrain
Practise →
No exam published for this title
NHRA publishes no exam for this title that we could find; check with NHRA
Not publishedNot publishedNot published
OMSB
Sultanate of Oman
Practise →
No exam published for this title
No OMSB exam published for this title; check with OMSB
Not publishedNot publishedNot published
Kuwait MOH
State of Kuwait
Practise →
No exam published for this title
Kuwait MOH publishes no exam for this title; check with the MOH
Not publishedNot publishedNot published

Highlighted rows publish figures for this exact Biomedical Engineering & Medical Devices exam; the others show the regulator's general rules.

What is on the Biomedical Engineering & Medical Devices exam: the full blueprint

The published blueprint splits into 12 domains and 67 testable sub-topics. For each one you can see its weight, how many questions this mock asks on it, and about how many the full bank holds.

  • Biomedical Equipment Technology~15%

    15 sub-topics · 4 in this mock · about 68 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Advanced Electronics and Electricity~15%

    15 sub-topics · 3 in this mock · about 68 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Anatomy and Physiology~10%

    10 sub-topics · 2 in this mock · about 45 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Mathematics~5%

    5 sub-topics · 2 in this mock · about 23 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Physics~5%

    5 sub-topics · 1 in this mock · about 23 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Chemistry~5%

    5 sub-topics · not in this mock · about 23 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Medical Informatics~10%

    10 sub-topics · 3 in this mock · about 45 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Professionalism~10%

    10 sub-topics · 1 in this mock · about 45 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Practical Experience~5%

    5 sub-topics · not in this mock · about 23 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Research and Design~5%

    5 sub-topics · 1 in this mock · about 23 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Patient Safety, Electrical Safety and Hazards~10%

    10 sub-topics · 6 in this mock · about 45 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

  • Regulation and Standards~5%

    5 sub-topics · 2 in this mock · about 23 in the full bank

    SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022

Bank counts are estimates from the published weights the bank is written to, rounded.

This mock samples 10 of 12 domains. The full bank covers all 12 across 450+ questions, including the 2 domains this mock does not reach.

See the full bank - AED 269

How to use your mock score

These bands are set against the 60% SCFHS pass mark and match the verdict on your result screen. On 25 questions, a score near the pass mark has a 95% margin of error of about ±19 points, so treat the band as a direction, not a forecast.

Below 55%

Below the line

On this sample you would likely fall short of the 60% pass mark today. Do not book yet. Start with your two weakest domains from the result, then sit full-length papers until you clear the line with room to spare.

55% to 69%

Close to the line

A few questions decide it. On a 150-question paper, every 1% is one or two questions, so drill your weakest domains and build a margin on timed papers before you book.

70% and above

On track

Clear of the pass mark on this sample. 25 questions is a small sample, so confirm it over full-length timed papers before you book: stamina and pacing fail candidates who know the material.

Getting an honest score

  1. 1Sit it cold, once. Do not look anything up. A diagnostic you help yourself through tells you nothing. The number you get is only useful if it is honest.
  2. 2Read every rationale, especially the ones you got right. Getting a question right by elimination is not the same as knowing it. The worked reasoning is free, whether or not you buy anything.
  3. 3Work on your weakest domain first. The breakdown orders your domains worst first for exactly this reason. Studying what you are already good at feels productive and moves nothing.
  4. 4Re-sit under full exam conditions later. The real paper is much longer than 25 questions and stamina matters. The full bank's 3 to 6 papers are written for exactly this.

A worked Biomedical Engineering & Medical Devices exam question

Every question in the mock and in the paid bank is written to this standard: a clinical stem, four plausible options and the reasoning for the answer, with its source.

Advanced Electronics and Electricity

An ECG shows strong 50 Hz noise on every lead in one room only. Which amplifier property mainly rejects this kind of mains interference picked up equally on both inputs?

  1. AA high output impedance in the final amplifier stage and its buffer
  2. BA low gain-bandwidth product in the following filter stage
  3. CA large input capacitance added on each electrode lead wire
  4. DA high common-mode rejection ratio in the differential input stageCorrect

Why: Mains pickup appears as a common-mode signal, and a differential amplifier with high CMRR cancels it while amplifying the difference between electrodes. Output impedance and gain-bandwidth do not reject common-mode noise, and extra input capacitance would degrade rather than help.

Source: Webster (ed.), Medical Instrumentation: Application and Design, 4th ed., 2010

Biomedical Engineering & Medical Devices sample questions with answers

10 more original exam-style Biomedical Engineering & Medical Devices questions, worked in full. They are not real or recalled exam questions, and none of them is in the mock above or on the bank page, so sit the mock first if you have not.

  1. Question 1 · Advanced Electronics and Electricity

    A diagnostic ECG has useful content up to 150 Hz and is digitised. Which sampling rate is the minimum that satisfies the Nyquist criterion?

    1. AMore than 300 samples per second
    2. B150 samples per second
    3. CMore than 75 samples per second
    4. D100 samples per second
    Show answer and explanation

    Correct answer: A. More than 300 samples per second

    The sampling rate must exceed twice the highest signal frequency, so above 300 Hz. Sampling at 150 or 100 Hz causes aliasing, and 75 Hz is half the bandwidth rather than double it.

    Reference: Webster (ed.), Medical Instrumentation: Application and Design, 4th ed., 2010

  2. Question 2 · Biomedical Equipment Technology

    The daily Bowie-Dick test on a pre-vacuum steam sterilizer shows an uneven colour change, with an unchanged patch in the centre of the sheet. What is the most likely cause?

    1. AThe chamber exceeded 134 degrees Celsius and stayed hot for too long a hold time
    2. BThe biological indicator vial was placed upside down inside the test pack
    3. CThe detergent used on the previous load was too alkaline and left a residue
    4. DIncomplete air removal, for example from a door gasket or vacuum leak
    Show answer and explanation

    Correct answer: D. Incomplete air removal, for example from a door gasket or vacuum leak

    A Bowie-Dick sheet tests air removal and steam penetration. A centre patch that did not change means trapped air, usually from a leak or a failing vacuum. Overheating would not leave a patch unchanged, and the biological indicator and detergent are not part of this test.

    Reference: AAMI ST79:2017, Comprehensive guide to steam sterilization

  3. Question 3 · Medical Informatics

    A networked infusion pump runs an unsupported operating system and the vendor cannot supply a patch. Which action best manages the risk?

    1. ADisable the pump alarms so that network traffic does not interrupt delivery
    2. BConnect it directly to the main hospital network so it can receive updates
    3. CLeave it unchanged because clinical devices are exempt from cybersecurity risk
    4. DPlace it on an isolated network segment with restricted firewall rules and document the residual risk
    Show answer and explanation

    Correct answer: D. Place it on an isolated network segment with restricted firewall rules and document the residual risk

    Where a device cannot be patched, network segmentation and access control reduce exposure, and the remaining risk is recorded under a risk management process. Disabling alarms harms patients, open connection raises risk, and no device is exempt.

    Reference: IEC 80001-1:2021, Application of risk management for IT networks incorporating medical devices

  4. Question 4 · Chemistry

    A flexible endoscope is being reprocessed. Why must manual cleaning with enzymatic detergent come before high-level disinfection?

    1. ADetergent sterilises the channels, so disinfection afterwards is a final rinse step
    2. BOrganic soil shelters microorganisms from the disinfectant and weakens its action
    3. CEnzymes raise the pH enough to kill spores that remain inside the lumen of the scope
    4. DSoil makes the disinfectant too concentrated for the scope material to tolerate safely
    Show answer and explanation

    Correct answer: B. Organic soil shelters microorganisms from the disinfectant and weakens its action

    Residual blood and tissue protect organisms and consume or neutralise chemical disinfectants, so cleaning must remove them first. Detergent does not sterilise, enzymes do not kill spores through pH, and soil dilutes rather than concentrates the disinfectant.

    Reference: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008

  5. Question 5 · Biomedical Equipment Technology

    How do the cut and coagulation modes of an electrosurgical unit differ?

    1. ACut uses an interrupted waveform, and coagulation uses a continuous low-voltage one
    2. BBoth use the same waveform at different foot pedal pressures
    3. CCut uses direct current, and coagulation uses alternating current
    4. DCut is continuous and low voltage, coagulation is pulsed and higher voltage
    Show answer and explanation

    Correct answer: D. Cut is continuous and low voltage, coagulation is pulsed and higher voltage

    A continuous waveform concentrates heat to vaporise tissue for cutting, while a pulsed, higher-voltage waveform gives time for cooling and produces coagulation. Reversing the modes is wrong, and the waveforms are set by the generator, not by pedal pressure. Both modes use high-frequency alternating current.

    Reference: Webster, Medical Instrumentation: Application and Design, 4th ed. (2010)

  6. Question 6 · Advanced Electronics and Electricity

    A step-down transformer has 1150 primary turns and changes 230 V to 12 V. How many secondary turns does it have?

    1. A60
    2. B30
    3. C120
    4. D600
    Show answer and explanation

    Correct answer: A. 60

    The voltage ratio equals the turns ratio, so the secondary turns are 1150 x 12 / 230 = 60. The values 30 and 120 are half and double that figure, and 600 comes from an incorrect ratio.

    Reference: Sedra and Smith, Microelectronic Circuits, 7th ed. (2014)

  7. Question 7 · Mathematics

    A blood pressure of 120 mmHg is converted to kilopascals. Which value is closest?

    1. A12 kPa
    2. B16 kPa
    3. C90 kPa
    4. D160 kPa
    Show answer and explanation

    Correct answer: B. 16 kPa

    One mmHg is about 0.133 kPa, so 120 mmHg is about 16 kPa. The value 12 kPa is about 90 mmHg. The values 90 and 160 come from using the wrong factor.

    Reference: NIST Guide for the Use of the International System of Units (2008)

  8. Question 8 · Medical Informatics

    Which backup arrangement follows the 3-2-1 rule for clinical data?

    1. AThree copies on one server in the same room
    2. BTwo copies on one disk, with a tape beside the server
    3. CThree copies on two media, one off site
    4. DOne copy on three media kept on site
    Show answer and explanation

    Correct answer: C. Three copies on two media, one off site

    The 3-2-1 rule keeps three copies, on two different media, with one copy away from the site so that a single event cannot destroy all of them. Copies on one server or in one room share the same risk.

    Reference: US-CERT/CISA Data Backup Options (2012)

  9. Question 9 · Practical Experience

    When calibrating a device, which accuracy ratio between the reference standard and the device under test is generally accepted as a minimum?

    1. AA ratio of at least 4 to 1
    2. B1 to 1, as both are calibrated
    3. C2 to 1 for general devices
    4. D10 to 1 for electrical safety tests
    Show answer and explanation

    Correct answer: A. A ratio of at least 4 to 1

    A reference that is at least four times more accurate than the device under test is the traditional test accuracy ratio and keeps the reference error from masking the device error. A 1 to 1 ratio leaves no margin, and a ratio of 2 to 1 is weaker than the usual rule.

    Reference: MIL-STD-45662A Calibration Systems Requirements (1988); ANSI/NCSL Z540.3 (2006)

  10. Question 10 · Patient Safety, Electrical Safety and Hazards

    In MRI safety zoning, which zone contains the scanner magnet?

    1. AZone I
    2. BZone II
    3. CZone III
    4. DZone IV
    Show answer and explanation

    Correct answer: D. Zone IV

    Zone IV is the magnet room, where the field is always on and access is supervised. Zone I is the public area, Zone II is reception and screening, and Zone III is the restricted control area.

    Reference: ACR Manual on MR Safety (2020)

Last reviewed September 2026. About these questions: every item is original, written to the published blueprint, with a rationale and a reference, and no recalled or leaked exam content is used. How we write our questions.

Free mock vs the full Biomedical Engineering & Medical Devices bank

The mock tells you where you stand. The bank is how you close the gap.

Free Biomedical Engineering & Medical Devices mock compared with the full bank
What you getFree mockFull bank
Questions25450+
Papers1 short timed set3 to 6 full-length papers, about 150 each
Blueprint coverage10 of 12 domainsAll 12 domains, 67 sub-topics
Worked rationaleEvery questionEvery question
ReferenceEvery questionEvery question
Timing30 minutes in your browserExam-length papers to sit against the clock
FormatOnline, freePDF sent to you, yours to keep
PriceFreeAED 269, paid once

Full Biomedical Engineering & Medical Devices bank

Close your gaps before exam day

  • 450+ original questions in 3 to 6 full-length papers
  • A worked rationale and a reference on every question
  • Prepared for your regulator's blueprint and pass mark

One-time price

AED 269

About a third of one exam fee (about AED 880 per attempt)

Unlock the full bank →

No subscription · yours to keep

Frequently asked questions

Is this Biomedical Engineering & Medical Devices mock test really free?
Yes. All 25 questions, your score, the domain breakdown and the full worked answers are free, with no sign-up, no email and no payment. The paid Biomedical Engineering & Medical Devices bank is a separate, much larger set of 450+ questions across 3 to 6 papers at AED 269.
Are these real Biomedical Engineering & Medical Devices exam questions?
No. They are original questions written to the published Biomedical Engineering & Medical Devices blueprint and drawn from our full bank, not leaked or recalled items. Recalled questions breach exam security rules and are often out of date. These follow the exam's format and domain weighting.
What is the pass mark for the Biomedical Engineering & Medical Devices exam?
By regulator: SCFHS (pass mark 60%), each from the regulator's own exam documents. Regulators not listed publish no figure for this exam, so confirm yours before booking. Most report only pass or fail and never release your score, which is why a diagnostic like this is useful.
How long is the Biomedical Engineering & Medical Devices exam and how many questions?
SCFHS: 150 MCQs in 180 minutes, from each regulator's own exam documents. Regulators not listed publish no length for this exam, so confirm theirs in the current candidate bulletin.
What topics does the Biomedical Engineering & Medical Devices exam cover?
The published blueprint has 12 domains and 67 testable sub-topics. The heaviest are Biomedical Equipment Technology (~15%), Advanced Electronics and Electricity (~15%) and Anatomy and Physiology (~10%). The full domain list with weights is on this page.
How is the full Biomedical Engineering & Medical Devices bank different from this mock?
This mock is 25 questions across 10 of the 12 blueprint domains, enough to rank your domains. The full bank is 450+ questions in 3 to 6 full-length papers of about 150 questions, covering every domain, with about 68 questions on Biomedical Equipment Technology alone. Every question has a worked rationale and a reference. It costs AED 269, paid once.
What score should I aim for on this Biomedical Engineering & Medical Devices mock?
Aim for at least 70%, which is 10 points over the 60% SCFHS pass mark. Below 55% you would likely fall short today; between 55% and 69% a few questions decide it. On 25 questions a score near the pass mark has a margin of error of about ±19 points, so read it as a direction.
Does passing this mock mean I will pass the real exam?
No, and we will not claim otherwise. 25 questions cannot predict a full-length paper: the margin of error is about ±19 points. What it does reliably show is which blueprint domains you are weakest in, so you know where to start.
Which regulators is this Biomedical Engineering & Medical Devices mock for?
It is written to the shared Biomedical Engineering & Medical Devices blueprint, so it fits SCFHS. SCFHS publishes an exam for this exact title. What changes by regulator is the length, the pass mark and the local regulation questions.
How many attempts do I get at the real exam?
In the UAE, three in total across DHA, DOH, MOHAP and SHA under the Unified Professional Qualification Requirements, not three per authority. Each attempt costs roughly AED 880, which is why it pays to find your weak domains first.

Other free mock exams in allied

From a 25-question mock to the full Biomedical Engineering & Medical Devices bank

450+ original questions in 3 to 6 full-length papers, each with a worked rationale and a reference, written to the same blueprint as the free mock on this page. One price: AED 269.