Central Sterile Services Technician Prometric exam questions with answers

15 original practice questions written to the Central Sterile Services Technician exam blueprint, each with the answer and why the other options are wrong. Below them: the Central Sterile Services Technician exam's format, pass mark and fee at DHA, DOH, SCFHS, QCHP and every other GCC regulator, from their own published rules.

15
Free questions
50%
DHA pass mark
150
Questions on the DHA exam
12
Questions in the free mock

Quick answer

The Central Sterile Services Technician exam is 150 MCQs in 3 hours at DHA (pass mark 50%). Below are 15 original practice questions written to the official blueprint, each with the answer and why the other options are wrong, plus a free 12-question timed mock.

The Central Sterile Services Technician exam in every GCC country

DHA publish an exact Central Sterile Services Technician exam; the others apply one format to every title. Figures come from each regulator's own exam pages - anything not published says so.

Central Sterile Services Technician licensing exam in every GCC country
RegulatorExamFormatPass markFee per attemptPractise
DHA
Dubai
Central Sterile Services (CSS) Technician (STE5652)150 MCQs in 3 hours50%USD 240 (about AED 880)DHA questions →
DOH
Abu Dhabi, Al Ain and Al Dhafra
DOH licensing exam
Regulator-wide format
Written + oral/OSCEPass / fail onlyNot publishedDOH questions →
MOHAP
Northern Emirates - Ajman, Umm Al Quwain, Ras Al Khaimah and Fujairah
MOHAP licensing exam
Regulator-wide format
Computer-basedNot publishedNot publishedMOHAP questions →
SHA
Emirate of Sharjah
SHA licensing exam
Regulator-wide format
Set at assessmentNot publishedNot publishedSHA questions →
SCFHS
Kingdom of Saudi Arabia
SCFHS licensing exam
Regulator-wide format
200 MCQs, 240 min500-560 of 800Not publishedSCFHS questions →
QCHP
State of Qatar
QCHP licensing exam
Regulator-wide format
150 MCQs, 3 hrs50-65%Not publishedQCHP questions →
NHRA
Kingdom of Bahrain
NHRA licensing exam
Regulator-wide format
Prometric CBTNot publishedNot publishedNHRA questions →
OMSB
Sultanate of Oman
OMSB licensing exam
Regulator-wide format
Computer-based OC examNot publishedNot publishedOMSB questions →
Kuwait MOH
State of Kuwait
Kuwait MOH licensing exam
Regulator-wide format
150 MCQs, 170 min60-70%Not publishedKuwait MOH questions →

exact exam published for Central Sterile Services Technician

Pass marks, attempts and fees change - confirm yours on the regulator's exam page before you book.

15 Central Sterile Services Technician exam questions with answers

Choose your answer first, then open the explanation. The set covers the main blueprint domains at a mix of easy, medium and hard.

  1. Question 1Decontamination: Cleaning and Disinfectionmedium

    How does an ultrasonic cleaner remove soil from instruments?

    1. ACavitation of tiny bubbles
    2. BHigh-pressure water jets
    3. CThermal coagulation of protein
    4. DChemical sterilisation
    Show answer and explanation

    Answer: A. Cavitation of tiny bubbles

    High-frequency sound waves create tiny bubbles that implode, a process called cavitation, dislodging soil from crevices and joints. It cleans but does not disinfect or sterilise.

  2. Question 2Decontamination: Cleaning and Disinfectionhard

    Why are blood-soiled instruments first rinsed in cool water rather than hot water?

    1. AHot water coagulates proteins onto surfaces
    2. BCool water sterilises the instruments
    3. CHot water damages enzymatic detergents only
    4. DCool water dries faster
    Show answer and explanation

    Answer: A. Hot water coagulates proteins onto surfaces

    Water above about 45 degrees Celsius coagulates blood proteins and fixes them to the instrument, making them harder to remove. Cool water keeps proteins soluble for the cleaning stage.

  3. Question 3Inspection, Assembly and Packagingmedium

    Where should a technician write on a paper-plastic peel pouch?

    1. AOn the paper side with a ballpoint pen
    2. BOn the plastic side or a label
    3. CAcross the heat seal
    4. DInside the pouch on a paper slip
    Show answer and explanation

    Answer: B. On the plastic side or a label

    Labelling is done on the plastic side or on a label so that ink does not penetrate or damage the paper barrier. Writing across seals can compromise their integrity.

  4. Question 4Low-Temperature Sterilisationmedium

    Which packaging material is unsuitable for hydrogen peroxide gas plasma sterilisation?

    1. ACellulose paper or linen
    2. BPolypropylene wrap
    3. CTyvek-plastic pouches
    4. DValidated rigid containers
    Show answer and explanation

    Answer: A. Cellulose paper or linen

    Cellulose materials such as paper and linen absorb hydrogen peroxide and can cause cycle failure. Polypropylene wraps and Tyvek pouches are compatible.

  5. Question 5Steam Sterilisationeasy

    How does saturated steam kill microorganisms?

    1. ABy oxidising cell walls chemically
    2. BBy denaturing and coagulating proteins
    3. CBy dehydrating spores slowly
    4. DBy damaging DNA with radiation
    Show answer and explanation

    Answer: B. By denaturing and coagulating proteins

    Moist heat transfers energy rapidly as steam condenses, causing irreversible denaturation and coagulation of microbial proteins. Dry heat works more slowly by oxidation.

  6. Question 6Microbiology and Infection Prevention Principlesmedium

    Which of the following is most resistant to chemical disinfectants and heat?

    1. AEnveloped viruses
    2. BMycobacteria
    3. CBacterial spores
    4. DVegetative bacteria
    Show answer and explanation

    Answer: C. Bacterial spores

    Apart from prions, bacterial spores are the most resistant to disinfection and sterilisation, followed by mycobacteria. Enveloped viruses such as HIV are among the easiest to inactivate.

  7. Question 7Sterile Storage and Distributionhard

    Which shelving arrangement is appropriate for sterile storage?

    1. AUnder the sink for easy access
    2. BOn the floor near the entrance door
    3. COn top shelves touching sprinklers
    4. DRaised off the floor, below sprinklers
    Show answer and explanation

    Answer: D. Raised off the floor, below sprinklers

    Sterile items are stored 8 to 10 inches (about 20 to 25 cm) above the floor and 18 inches (about 45 cm) below the ceiling or sprinklers, away from water sources. Floor and under-sink storage risk contamination.

  8. Question 8Workplace Safety and Professional Conducteasy

    What personal protective equipment is required when manually cleaning instruments in the decontamination area?

    1. ADisposable examination gloves only
    2. BFluid-resistant gown, gloves, face shield
    3. CCotton apron, cap and shoe covers
    4. DSterile gown with sterile gloves only
    Show answer and explanation

    Answer: B. Fluid-resistant gown, gloves, face shield

    Manual cleaning creates splashes, so staff wear a fluid-resistant gown, heavy-duty gloves, and eye and face protection such as goggles with a mask or a face shield. Gloves alone leave the face and body exposed.

Halfway - how are you scoring?

Test yourself under real exam conditions

The free 12-question mock is timed and scored against the pass mark, domain by domain, so you see exactly where you are losing marks. The full Central Sterile Services Technician bank has 3 full-length papers (about 450 questions) for AED 179, one-time.

  1. Question 9Monitoring, Quality Assurance and Traceabilitymedium

    A steam load contains orthopaedic implants. When may the load be released for use?

    1. AAs soon as the cycle ends
    2. BAfter the BI result is negative
    3. CWhen the external tape changes
    4. DAfter the load has cooled
    Show answer and explanation

    Answer: B. After the BI result is negative

    Loads containing implants should be monitored with a biological indicator and quarantined until the result is negative. Early release is limited to documented emergencies.

  2. Question 10Steam Sterilisationmedium

    A wrapped tray is found to be wet after the cycle has ended and cooled. What should be done?

    1. ATreat it as unsterile and reprocess
    2. BDry it in a warming cabinet and use
    3. CUse it if the indicator has changed
    4. DOpen it and wipe the moisture
    Show answer and explanation

    Answer: A. Treat it as unsterile and reprocess

    Moisture can wick microorganisms through the packaging, so wet packs are considered contaminated and must be reprocessed, and the cause investigated. A changed indicator does not overcome this.

  3. Question 11Steam Sterilisationmedium

    How should peel pouches be placed in a steam steriliser?

    1. AOn edge, paper facing plastic
    2. BFlat and stacked on each other
    3. CUnder heavy metal trays
    4. DPacked tightly to fill the chamber
    Show answer and explanation

    Answer: A. On edge, paper facing plastic

    Pouches are placed on edge in a rack, paper side to plastic side, so steam can penetrate and condensate can drain. Stacking flat or overloading traps air and causes wet packs.

  4. Question 12Steam Sterilisationmedium

    A load has just been removed from the steriliser. How should it be handled?

    1. ACool quickly under an air vent
    2. BPlace on a cold metal surface
    3. CStore immediately in closed cabinets
    4. DCool on the cart away from vents
    Show answer and explanation

    Answer: D. Cool on the cart away from vents

    Loads should cool on the cart in a low-traffic area away from air vents, untouched, to prevent condensation forming inside packs. Cold surfaces and air currents cause condensation and possible contamination.

  5. Question 13Steam Sterilisationmedium

    The morning Bowie-Dick test sheet shows uneven colour change with a pale centre. What is the correct action?

    1. ARepeat the test after the first load
    2. BRun patient loads but add extra indicators
    3. CWithdraw the steriliser and report it
    4. DUse the steriliser for gravity cycles only
    Show answer and explanation

    Answer: C. Withdraw the steriliser and report it

    A pale centre indicates inadequate air removal or air leak, so the steriliser must not be used for porous loads until fixed and retested. Running patient loads risks non-sterile items.

  6. Question 14Sterile Storage and Distributionmedium

    A department uses event-related sterility. What determines whether a stored pack is still sterile?

    1. AThe number of times it was moved
    2. BA fixed 30-day expiry for all packs
    3. CThe colour of the internal indicator
    4. DWhether the packaging is intact and dry
    Show answer and explanation

    Answer: D. Whether the packaging is intact and dry

    Under event-related sterility, a pack remains sterile unless an event such as tearing, wetting or crushing compromises the barrier. Handling should be minimised, but time alone does not make it unsterile.

  7. Question 15Workplace Safety and Professional Conductmedium

    A technician opening a returned tray finds a loose scalpel blade among the instruments. What is the safest approach?

    1. ALift it out with an instrument
    2. BPick it out carefully by hand
    3. CLeave it in for the washer
    4. DReturn the tray unopened
    Show answer and explanation

    Answer: A. Lift it out with an instrument

    Loose sharps should be removed with an instrument, never fingers, placed in a sharps container, and the incident reported so theatre practice can improve. Leaving it risks injury to the next handler.

What the Central Sterile Services Technician exam covers

The blueprint groups questions into these domains. Weight your revision the same way - the heavier domains carry more of your score.

Microbiology and Infection Prevention Principles

~10%

Chain of infection and standard precautions · Spaulding classification · Hierarchy of microbial resistance to disinfection and sterilisation

Decontamination: Cleaning and Disinfection

~18%

Manual cleaning · Detergents · Ultrasonic cleaning and cavitation

Inspection, Assembly and Packaging

~15%

Inspection with lighted magnification and function testing · Insulation testing of laparoscopic instruments · Instrument lubrication and care

Steam Sterilisation

~18%

Principles of moist heat sterilisation · Gravity displacement vs pre-vacuum cycles · Time and temperature parameters

Low-Temperature Sterilisation

~10%

Ethylene oxide · Hydrogen peroxide gas plasma and vapour systems · Material and packaging compatibility

Monitoring, Quality Assurance and Traceability

~14%

Physical monitoring · Chemical indicator types and classes · Biological indicators and process challenge devices

Sterile Storage and Distribution

~7%

Event-related sterility and shelf life · Storage conditions · Stock rotation and inventory management

Workplace Safety and Professional Conduct

~8%

Personal protective equipment in the decontamination area · Sharps safety and injury reporting · Chemical safety

How to answer these questions

1

Questions test applied practice: what you would do with this patient, this result or this image - not textbook definitions.

2

Safety items (radiation, infection control, patient identification, equipment checks) are high-yield and quick to revise.

3

Under time pressure, flag and move on: every question carries the same mark.

4

Aim to score comfortably above your regulator's pass mark - 10 to 15 points of margin - on timed, full-length practice before you book.

Central Sterile Services Technician exam questions: FAQs

How many questions are in the Central Sterile Services Technician Prometric exam?
It depends on the regulator: DHA 150 MCQs in 3 hours. The full table above lists every GCC regulator.
What is the pass mark for the Central Sterile Services Technician exam?
DHA: 50%. Pass marks are set exam by exam, so use your own regulator's figure.
Are these real exam questions?
No. Regulators do not release their papers, and anything sold as a leaked paper is unofficial and risky to rely on. These are original questions written to the official exam blueprint, at the level and in the style of the real exam, each with a worked explanation.
Is the Central Sterile Services Technician exam the same in every GCC country?
The clinical content is broadly similar because the exams test the same safe practice, but the format, length, pass mark, attempts and fees differ by regulator. Prepare on the content once, then check your regulator's exact rules before you book.
How should I use these questions?
Answer each one before opening the explanation, and note why each wrong option is wrong. Then take the free 12-question timed mock to see your score against the pass mark, and move to full-length papers when you score comfortably above it.

Last reviewed September 2026. Questions are original and written to the published exam blueprint; they are not taken from any real paper.

Pass the Central Sterile Services Technician exam first time

Not sure which regulator's exam you need, or whether you are exempt? Tell us your profession and target country - we confirm your route and the exact exam before you book.