Biomedical Engineering & Medical Devices Prometric exam questions with answers
15 original practice questions written to the Biomedical Engineering & Medical Devices exam blueprint, each with the answer and why the other options are wrong. Below them: the Biomedical Engineering & Medical Devices exam's format, pass mark and fee at DHA, DOH, MOHAP, SHA, SCFHS, QCHP, NHRA, OMSB and Kuwait MOH, from their own published rules.
- 15
- Free questions
- 60%
- SCFHS pass mark
- 150
- Questions on the SCFHS exam
- 25
- Questions in the free mock
Quick answer
The Biomedical Engineering & Medical Devices exam is 150 MCQs in 180 minutes at SCFHS (pass mark 60%). Below are 15 original practice questions written to the official blueprint, each with the answer and why the other options are wrong, plus a free 25-question timed mock.
The full Biomedical Engineering & Medical Devices bank: from 3 papers, ~150 questions each
AED 269 one-time
Every question with a rationale and a reference. PDF, lifetime access.
Tell us your authority and we send the bank for that exam, usually the same day.
The Biomedical Engineering & Medical Devices exam in every GCC country
SCFHS publish an exact Biomedical Engineering & Medical Devices exam; the others apply one format to every title. Figures come from each regulator's own exam pages - anything not published says so.
| Regulator | Exam | Format | Pass mark | Fee per attempt |
|---|---|---|---|---|
Saudi Classification Examination - Biomedical Engineering (Medical Devices) Specialist | 150 MCQs in 180 minutes | 60% | Not published | |
No exam published for this title Not in the DHA CBT list; DHA's guideline assesses licensed titles that are not listed by oral | Not published | Not published | Not published | |
No exam published for this title DOH publishes no exam for this title that we could find; check with DOH | Not published | Not published | Not published | |
No exam published for this title MOHAP publishes no exam for this title that we could find; check with MOHAP | Not published | Not published | Not published | |
No exam published for this title SHA publishes no exam for this title that we could find; check with SHA | Not published | Not published | Not published | |
No exam published for this title No DHP Qatar exam blueprint published for this title; check with DHP | Not published | Not published | Not published | |
No exam published for this title NHRA publishes no exam for this title that we could find; check with NHRA | Not published | Not published | Not published | |
No exam published for this title No OMSB exam published for this title; check with OMSB | Not published | Not published | Not published | |
No exam published for this title Kuwait MOH publishes no exam for this title; check with the MOH | Not published | Not published | Not published |
Highlighted rows publish figures for this exact Biomedical Engineering & Medical Devices exam; the others show the regulator's general rules.
Pass marks, attempts and fees change - confirm yours on the regulator's exam page before you book.
15 Biomedical Engineering & Medical Devices exam questions with answers
Choose your answer first, then open the explanation. The set covers the main blueprint domains at a mix of easy, medium and hard.
- Question 1Advanced Electronics and Electricitymedium
An inverting op-amp amplifier has a 10 kilohm input resistor and a 100 kilohm feedback resistor. A 0.2 V signal is applied. What is the output voltage?
- APlus 2 V
- BMinus 2 V
- CMinus 0.02 V
- DPlus 2.2 V
Show answer and explanation
Answer: B. Minus 2 V
Inverting gain is minus Rf over Rin, which is minus 10, so 0.2 V gives minus 2 V. Plus 2 V ignores the inversion, minus 0.02 V divides instead of multiplying, and plus 2.2 V applies the non-inverting gain of 11.
Reference: Horowitz and Hill, The Art of Electronics, 3rd ed., 2015
- Question 2Anatomy and Physiologyhard
A finger pulse oximeter reads 98% on a patient who is confused after smoke inhalation, yet the arterial blood is poorly oxygenated. What explains the misleading reading?
- AMethaemoglobin has no absorbance at either wavelength, so it is ignored
- BCarboxyhaemoglobin absorbs red light much like oxyhaemoglobin, so it is counted as saturated
- CFetal haemoglobin blocks infrared light and raises the ratio
- DLow skin temperature raises the reading by about ten points on most devices
Show answer and explanation
Answer: B. Carboxyhaemoglobin absorbs red light much like oxyhaemoglobin, so it is counted as saturated
Two-wavelength oximeters cannot tell carboxyhaemoglobin from oxyhaemoglobin, so carbon monoxide poisoning gives a falsely normal reading. Methaemoglobin does absorb light, fetal haemoglobin is not the cause here, and cold skin gives poor signal rather than a fixed overread.
Reference: Webster (ed.), Medical Instrumentation: Application and Design, 4th ed., 2010
- Question 3Biomedical Equipment Technologymedium
An oscillometric blood pressure monitor shows an unusual diastolic value on one patient. Which pressure does the device detect directly from the cuff oscillations?
- AThe systolic pressure at the first Korotkoff sound heard by the cuff microphone
- BThe mean arterial pressure, found at the maximum oscillation amplitude
- CThe diastolic pressure at the last audible sound under the cuff
- DThe pulse pressure measured as the width of each oscillation envelope
Show answer and explanation
Answer: B. The mean arterial pressure, found at the maximum oscillation amplitude
Oscillometric devices find the cuff pressure with the largest oscillations, which approximates mean arterial pressure, and then calculate systolic and diastolic values with a manufacturer algorithm. Korotkoff sounds belong to auscultation and are not sensed, and pulse pressure is not read from oscillation width.
Reference: ISO 81060-2:2018, Non-invasive sphygmomanometers, clinical validation
- Question 4Chemistrymedium
A blood gas analyser measures partial pressure of oxygen with a Clark electrode. What produces the signal?
- AOxygen oxidises a silver anode, producing a voltage that is proportional to blood pH
- BOxygen is reduced at a platinum cathode, giving a current that follows PO2
- CHydrogen ions cross a thin glass membrane and generate a Nernst potential at the sensor
- DLight absorbed by oxygen at 660 nm changes the resistance of a photosensitive cell
Show answer and explanation
Answer: B. Oxygen is reduced at a platinum cathode, giving a current that follows PO2
The Clark electrode reduces oxygen diffusing through a membrane at a polarised platinum cathode, and the current follows PO2. Glass membrane potentials describe the pH electrode, and optical absorbance is how oximeters work, not the Clark cell.
Reference: Khandpur, Handbook of Biomedical Instrumentation, 3rd ed., 2014
- Question 5Medical Informaticseasy
A CT scanner sends images to the hospital archive. Which standard defines the image format and network transfer between the modality and the archive?
- AHL7 version 2 ADT
- BICD-10
- CDICOM
- DLOINC
Show answer and explanation
Answer: C. DICOM
DICOM defines how medical images and related data are formatted and exchanged between modalities and archives. HL7 ADT carries admission messages, ICD-10 codes diagnoses and LOINC codes laboratory observations.
Reference: NEMA, DICOM PS3, current edition
- Question 6Medical Informaticsmedium
A vendor describes its clinical data interface as based on resources exchanged through a web API in JSON. Which standard is this?
- AHL7 version 2 pipe-delimited messaging
- BDICOM Part 10 files
- CHL7 FHIR
- DIEEE 11073 device communication
Show answer and explanation
Answer: C. HL7 FHIR
FHIR defines resources exchanged over RESTful web services in JSON or XML. HL7 v2 uses pipe-delimited segments, DICOM Part 10 is a file format for images, and IEEE 11073 covers point-of-care device data.
Reference: HL7 International, FHIR Release 4 (v4.0.1), 2019
- Question 7Research and Designhard
A study compares a new tympanic thermometer with a reference thermometer on 80 patients. Which analysis best shows agreement between the two?
- AA Bland-Altman plot of differences against means with limits of agreement
- BThe correlation coefficient between the two sets of readings taken on each patient
- CA comparison of the highest reading obtained from each device across the whole group
- DA count of how many readings are shown in the same colour on the device screen
Show answer and explanation
Answer: A. A Bland-Altman plot of differences against means with limits of agreement
Bland-Altman analysis shows bias and the range within which most differences fall. High correlation can coexist with poor agreement, and maximum values or display colour say nothing about agreement.
Reference: Bland and Altman, Lancet 1986;1:307-310
- Question 8Physicseasy
A transducer is needed for imaging the thyroid, a structure only a few centimetres below the skin. Which probe is the best choice?
- AA curvilinear array operating at about 2 to 5 MHz
- BA phased array operating at about 1 to 3 MHz
- CA linear array operating at about 10 to 15 MHz
- DA continuous-wave pencil probe at about 20 kHz
Show answer and explanation
Answer: C. A linear array operating at about 10 to 15 MHz
High frequency gives fine resolution at the cost of penetration, which suits superficial structures. Low-frequency probes penetrate deeper but resolve less, and a kilohertz continuous-wave probe is not an imaging transducer.
Reference: Bushberg et al., The Essential Physics of Medical Imaging, 4th ed., 2021
Halfway - how are you scoring?
Test yourself under real exam conditions
The free 25-question mock is timed and scored against the pass mark, domain by domain, so you see exactly where you are losing marks. The full Biomedical Engineering & Medical Devices bank starts at 3 full-length papers of about 150 questions each for AED 269, one-time, and you can add papers up to 6.
Tell us your authority and we send the bank for that exam, usually the same day.
- Question 9Physicseasy
A point X-ray source gives 4 mGy per hour at 1 m. Ignoring scatter, what is the rate at 2 m?
- A2 mGy per hour
- B0.5 mGy per hour
- C8 mGy per hour
- D1 mGy per hour
Show answer and explanation
Answer: D. 1 mGy per hour
Intensity falls with the square of distance, so doubling the distance divides it by four. A halving rate would be 2 mGy, 0.5 mGy divides by eight, and 8 mGy is an increase.
Reference: Bushberg et al., The Essential Physics of Medical Imaging, 4th ed., 2021
- Question 10Practical Experienceeasy
A bedside suction unit arrives with no display and no motor sound. Which step comes first in systematic troubleshooting?
- ACheck the power outlet, mains cable, fuse and power switch for faults
- BReplace the suction motor first and then test the unit again under load
- COpen the main board and re-solder the connectors that look doubtful
- DSend the unit for external calibration before the simple checks are made
Show answer and explanation
Answer: A. Check the power outlet, mains cable, fuse and power switch for faults
Start with the simplest and cheapest causes, such as supply, cable and fuse, before opening or replacing parts. Replacing the motor or reworking the board is premature and calibration does not fix a dead unit.
Reference: Dyro (ed.), Clinical Engineering Handbook, 2004
- Question 11Professionalismeasy
A nurse asks the technician to skip a failed alarm self-test on an infusion pump because the ward is short of pumps. What is the correct response?
- ADecline, tag the pump out of service and report it to the supervisor promptly
- BSkip the test this one time and record the omission in the log later in the shift
- CDisable the alarm so that the pump does not warn the nurse during the infusion
- DReturn it to the ward once the nurse signs a waiver accepting the risk
Show answer and explanation
Answer: A. Decline, tag the pump out of service and report it to the supervisor promptly
A pump with a failed alarm self-test is unsafe and must be removed from use until repaired. Skipping the test, disabling the alarm, or relying on a waiver from a nurse all leave patients exposed to unsafe equipment.
Reference: Dyro (ed.), Clinical Engineering Handbook, 2004
- Question 12Patient Safety, Electrical Safety and Hazardsmedium
Ward nurses complain about too many nuisance alarms on bedside monitors and want them turned off. What is the appropriate action?
- AReview and tailor alarm limits and settings through the hospital's alarm management process
- BTurn off the alarms on the monitors in the ward
- CMute the speakers permanently with tape
- DReplace the monitors with newer models straight away
Show answer and explanation
Answer: A. Review and tailor alarm limits and settings through the hospital's alarm management process
Alarm fatigue is addressed by clinically agreed limits, appropriate defaults and training, not by removing alarms. Disabling or muting them hides real danger, and buying new monitors does not fix settings.
Reference: IEC 60601-1-8:2006+A1:2012+A2:2020; Joint Commission NPSG.06.01.01
- Question 13Professionalismeasy
A technician finishes a preventive maintenance visit on a patient monitor. Which entry is essential in the maintenance record?
- AThe technician's signature and the date of the visit, kept on a separate sheet
- BThe purchase price of the monitor and the discount given by the vendor in the contract
- CThe asset number, date, tests with results, parts used and the outcome
- DA general comment that the monitor appeared to work well during the visit
Show answer and explanation
Answer: C. The asset number, date, tests with results, parts used and the outcome
A complete record lets later staff trace the device and prove tests were done. A signature alone, purchase details or a vague comment do not demonstrate what was tested or found.
Reference: IEC 62353:2014, Recurrent test and test after repair of medical electrical equipment
- Question 14Practical Experiencemedium
Which tasks belong in the scheduled preventive maintenance of a manual defibrillator?
- AReplacing the thermal paper roll and wiping the outside of the case clean
- BUpdating the hospital's insurance details for the device each financial year
- CRecalibrating the audio volume of the alarm tones and the charge tone
- DBattery capacity test, energy delivery test and a check of paddles and leads
Show answer and explanation
Answer: D. Battery capacity test, energy delivery test and a check of paddles and leads
Defibrillator readiness depends on battery capacity and accurate energy delivery, together with intact leads and paddles. Paper and cleaning alone do not test function, and the other items are unrelated.
Reference: IEC 62353:2014; IEC 60601-2-4:2010
- Question 15Regulation and Standardseasy
In Saudi Arabia, which authority regulates the marketing and post-market surveillance of medical devices?
- AThe Saudi Food and Drug Authority
- BThe Saudi Central Board for Accreditation of Healthcare Institutions
- CThe Ministry of Municipal and Rural Affairs
- DThe Saudi Commission for Health Specialties
Show answer and explanation
Answer: A. The Saudi Food and Drug Authority
SFDA regulates medical devices, including marketing authorisation and adverse event reporting. CBAHI accredits healthcare facilities, SCFHS registers and classifies practitioners, and the municipal ministry does not regulate devices.
Reference: SFDA, Medical Devices Interim Regulation (MDS-IR), current edition
What the Biomedical Engineering & Medical Devices exam covers
The blueprint groups questions into these domains. Weight your revision the same way - the heavier domains carry more of your score.
Biomedical Equipment Technology
~15%Ventilators · Defibrillators and monitors · Infusion and syringe pumps
Advanced Electronics and Electricity
~15%DC and AC circuit analysis · Operational amplifiers · Differential amplifiers, common-mode rejection and noise
Anatomy and Physiology
~10%Cardiac conduction and ECG waveforms · Blood pressure, mean arterial pressure and circulation · Respiratory mechanics, gas exchange and capnography
Mathematics
~5%Algebra, ratios and unit conversion · Logarithms and decibels · Flow, volume and dosing rate calculations
Physics
~5%Ultrasound · Ionising radiation · Magnetic resonance
Chemistry
~5%Disinfectants, detergents and the importance of cleaning before disinfection · Electrochemical sensors · Battery chemistries and corrosion
Medical Informatics
~10%DICOM, PACS and the modality worklist · HL7 version 2 and FHIR interoperability · Networked devices and IT risk management
Professionalism
~10%Ethics, confidentiality and protection of patient data · Refusing unsafe requests and escalating faults · Maintenance records and documentation
Practical Experience
~5%Systematic troubleshooting from power supply to subsystem · Defibrillator and monitor preventive maintenance · Traceable calibration and test uncertainty ratio
Research and Design
~5%Design controls · Method comparison and Bland-Altman analysis · Technology assessment and procurement
Patient Safety, Electrical Safety and Hazards
~10%Leakage current limits for B, BF and CF applied parts · Macroshock, microshock and grounding · Electrosurgical burns and return electrode contact
Regulation and Standards
~5%SFDA regulation of medical devices in Saudi Arabia · IEC 60601 family and IEC 62353 in-service testing · ISO 14971 risk management and its control hierarchy
How to answer these questions
Questions test applied practice: what you would do with this patient, this result or this image - not textbook definitions.
Safety items (radiation, infection control, patient identification, equipment checks) are high-yield and quick to revise.
Under time pressure, flag and move on: every question carries the same mark.
Aim to score comfortably above your regulator's pass mark - 10 to 15 points of margin - on timed, full-length practice before you book.
The full Biomedical Engineering & Medical Devices bank: from 3 papers, ~150 questions each
AED 269 one-time
Every question with a rationale and a reference. PDF, lifetime access.
Tell us your authority and we send the bank for that exam, usually the same day.
Biomedical Engineering & Medical Devices exam questions: FAQs
How many questions are in the Biomedical Engineering & Medical Devices Prometric exam?
What is the pass mark for the Biomedical Engineering & Medical Devices exam?
Are these real exam questions?
Is the Biomedical Engineering & Medical Devices exam the same in every GCC country?
How should I use these questions?
Last reviewed September 2026. Questions are original and written to the published exam blueprint; they are not taken from any real paper.