The Biomedical Engineering & Medical Devices exam at a glance
SCFHS (Saudi Arabia) publishes a Biomedical Engineering & Medical Devices exam of 150 MCQs in 180 minutes with a 60% pass mark. DHA, DOH, MOHAP, SHA, QCHP, NHRA, OMSB, Kuwait MOH publish no exam for this title.
- SCFHS · Saudi Arabia
- Saudi Classification Examination - Biomedical Engineering (Medical Devices) Specialist
- 150 MCQs in 180 minutes · pass mark 60% · checked October 2026
From each regulator's own exam documents. Every other regulator is in the full GCC table.
Try two questions from this bank
Original questions from the bank - pick an answer to see the rationale and reference.
A 12 V supply is connected across a 120 ohm resistor and a 180 ohm resistor in series. What current flows?
Choose an answer to see the rationale.
An ICU ventilator keeps alarming for high airway pressure. The patient looks comfortable, the circuit is dry and the set pressure limit is unchanged. Which fault should the technician check first?
Choose an answer to see the rationale.
Not sure where you stand? Sit 25 of these questions free, timed at real exam pace, and see which domains you are weakest in.
Take the free mock exam →Liked those? The full bank has 450 questions at this standard - every one with a rationale and a reference.
Get the full bank - AED 269What's inside
- ✓3 to 6 full-length papers (~150 Q each)
- ✓No question repeated between papers
- ✓Four options, one unambiguous best answer
- ✓A full rationale on every question
- ✓A real guideline or textbook reference
- ✓Every sub-topic in the published blueprint
- ✓Researched by Neelim's exam research team
The Biomedical Engineering & Medical Devices exam in every GCC country
| Regulator | Exam | Format | Pass mark | Fee per attempt |
|---|---|---|---|---|
Saudi Classification Examination - Biomedical Engineering (Medical Devices) Specialist | 150 MCQs in 180 minutes | 60% | Not published | |
No exam published for this title Not in the DHA CBT list; DHA's guideline assesses licensed titles that are not listed by oral | Not published | Not published | Not published | |
No exam published for this title DOH publishes no exam for this title that we could find; check with DOH | Not published | Not published | Not published | |
No exam published for this title MOHAP publishes no exam for this title that we could find; check with MOHAP | Not published | Not published | Not published | |
No exam published for this title SHA publishes no exam for this title that we could find; check with SHA | Not published | Not published | Not published | |
No exam published for this title No DHP Qatar exam blueprint published for this title; check with DHP | Not published | Not published | Not published | |
No exam published for this title NHRA publishes no exam for this title that we could find; check with NHRA | Not published | Not published | Not published | |
No exam published for this title No OMSB exam published for this title; check with OMSB | Not published | Not published | Not published | |
No exam published for this title Kuwait MOH publishes no exam for this title; check with the MOH | Not published | Not published | Not published |
Highlighted rows publish figures for this exact Biomedical Engineering & Medical Devices exam; the others show the regulator's general rules.
From each regulator's own exam pages; anything a regulator does not publish says so. Assessment mode also depends on your licence title, so confirm your route on your eligibility notice.
Try 15 free Biomedical Engineering & Medical Devices questions with answersDifferent from the samples on this page - each with a full explanationStart →What the Biomedical Engineering & Medical Devices exam covers
The published blueprint for this exam breaks into 12 domains and 67 testable sub-topics. Our bank covers every one of them - here is the full map, so you can see exactly what you are expected to know.
Biomedical Equipment Technology~15% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)15 topics
- Ventilators: modes, alarms, circuit faults and pressure and flow sensing
- Defibrillators and monitors: energy delivery testing and ECG, SpO2, NIBP and capnography principles
- Infusion and syringe pumps: occlusion detection and flow accuracy testing
- Electrosurgical units, ultrasound and imaging equipment basics
- Sterilisers and washer-disinfectors: steam cycles, Bowie-Dick and indicators
- Dialysis, anaesthesia machine and incubator fundamentals
- Calibration and performance verification with test equipment
- Preventive maintenance, corrective repair and failure analysis
- +7 more in the bank
Advanced Electronics and Electricity~15% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)15 topics
- DC and AC circuit analysis: Ohm's law, series and parallel networks
- Operational amplifiers: inverting and non-inverting gain
- Differential amplifiers, common-mode rejection and noise
- Filters, cut-off frequency and signal conditioning
- Analogue-to-digital conversion and the Nyquist criterion
- Power supplies, batteries and switch-mode converters
- Transducers and sensors: strain gauge, thermistor and piezoelectric
- Digital logic and microcontroller basics
- +7 more in the bank
Anatomy and Physiology~10% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)10 topics
- Cardiac conduction and ECG waveforms
- Blood pressure, mean arterial pressure and circulation
- Respiratory mechanics, gas exchange and capnography
- Blood oxygen transport and haemoglobin variants
- Renal function and the principles of dialysis
- Nervous system and bioelectric signals
- +4 more in the bank
Mathematics~5% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)5 topics
- Algebra, ratios and unit conversion
- Logarithms and decibels
- Flow, volume and dosing rate calculations
- Statistics for measurement error and calibration
- +1 more in the bank
Physics~5% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)5 topics
- Ultrasound: frequency, resolution and penetration
- Ionising radiation: inverse square law and shielding
- Magnetic resonance: Larmor frequency and field strength
- Heat, pressure and fluid flow in devices
- +1 more in the bank
Chemistry~5% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)5 topics
- Disinfectants, detergents and the importance of cleaning before disinfection
- Electrochemical sensors: Clark and pH electrodes
- Battery chemistries and corrosion
- Materials and biocompatibility
- +1 more in the bank
Medical Informatics~10% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)10 topics
- DICOM, PACS and the modality worklist
- HL7 version 2 and FHIR interoperability
- Networked devices and IT risk management
- Medical device cybersecurity and patching
- Electronic health record integration
- Data backup, retention and privacy
- +4 more in the bank
Professionalism~10% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)10 topics
- Ethics, confidentiality and protection of patient data
- Refusing unsafe requests and escalating faults
- Maintenance records and documentation
- Conflicts of interest and gifts
- Communication with clinicians and vendors
- Continuing professional development and scope of practice
- +4 more in the bank
Practical Experience~5% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)5 topics
- Systematic troubleshooting from power supply to subsystem
- Defibrillator and monitor preventive maintenance
- Traceable calibration and test uncertainty ratio
- Spare parts, vendor service and equipment lifecycle
- +1 more in the bank
Research and Design~5% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)5 topics
- Design controls: verification and validation
- Method comparison and Bland-Altman analysis
- Technology assessment and procurement
- Prototyping, usability and human factors
- +1 more in the bank
Patient Safety, Electrical Safety and Hazards~10% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)10 topics
- Leakage current limits for B, BF and CF applied parts
- Macroshock, microshock and grounding
- Electrosurgical burns and return electrode contact
- Outlet and ground testing in patient areas
- MRI safety labels: MR Safe, MR Conditional and MR Unsafe
- Alarm management and alarm fatigue
- Fire, oxygen and electrical hazards
- Infection control and device reprocessing
- +2 more in the bank
Regulation and Standards~5% (SCFHS Biomedical Engineering (Medical Devices) Specialist blueprint, 6719, 18 July 2022)5 topics
- SFDA regulation of medical devices in Saudi Arabia
- IEC 60601 family and IEC 62353 in-service testing
- ISO 14971 risk management and its control hierarchy
- ISO 13485 quality management
- CBAHI accreditation requirements for equipment management
12 worked Biomedical Engineering & Medical Devices practice questions
Original exam-style questions from the bank, spread across the blueprint domains. They are not real or recalled exam questions. Try each one, then open the answer for the rationale and reference. None of them appear in the free mock exam, so reading them will not spoil it.
Question 1 · Advanced Electronics and Electricity
A first-order RC low-pass filter uses a 1.6 kilohm resistor and a 0.1 microfarad capacitor. What is the approximate cut-off frequency?
- AAbout 100 Hz
- BAbout 1 kHz
- CAbout 10 kHz
- DAbout 6.3 kHz
Show answer and explanation
Correct answer: B. About 1 kHz
fc = 1 / (2 pi R C) = 1 / (2 x 3.14 x 1600 x 0.0000001), which is about 995 Hz. 100 Hz and 10 kHz are off by a factor of ten, and 6.3 kHz is 2 pi x 1 kHz, which confuses angular frequency with frequency in hertz.
Reference: Horowitz and Hill, The Art of Electronics, 3rd ed., 2015
Question 2 · Anatomy and Physiology
A technician is calibrating an ECG simulator. Which ECG component represents depolarisation of the ventricles?
- AThe P wave
- BThe QRS complex
- CThe T wave
- DThe PR segment
Show answer and explanation
Correct answer: B. The QRS complex
Ventricular depolarisation produces the QRS complex. The P wave is atrial depolarisation, the T wave is ventricular repolarisation, and the PR segment is the delay at the AV node.
Reference: Guyton and Hall, Textbook of Medical Physiology, 14th ed., 2020
Question 3 · Anatomy and Physiology
A patient monitor displays a blood pressure of 120/90 mmHg. What is the approximate mean arterial pressure?
- A100 mmHg
- B105 mmHg
- C90 mmHg
- D110 mmHg
Show answer and explanation
Correct answer: A. 100 mmHg
MAP is diastolic plus one third of the pulse pressure: 90 + (30 / 3) = 100 mmHg. 105 and 110 mmHg overweight the systolic value, and 90 mmHg is only the diastolic pressure.
Reference: Guyton and Hall, Textbook of Medical Physiology, 14th ed., 2020
Question 4 · Patient Safety, Electrical Safety and Hazards
Which patient leakage current limit in normal condition applies to a type CF applied part under the general standard for medical electrical equipment?
- A100 microamperes
- B500 microamperes
- C10 microamperes
- D5 milliamperes
Show answer and explanation
Correct answer: C. 10 microamperes
Type CF applied parts, used directly on the heart, are limited to 10 microamperes in normal condition and 50 in single fault. 100 and 500 microamperes are the type BF limits, and 5 milliamperes is the earth leakage limit.
Reference: IEC 60601-1:2005+AMD1:2012+AMD2:2020
Question 5 · Physics
Hydrogen protons precess at about 42.58 MHz per tesla. What is the approximate Larmor frequency on a 3 T MRI scanner?
- AAbout 64 MHz
- BAbout 43 MHz
- CAbout 128 MHz
- DAbout 255 MHz
Show answer and explanation
Correct answer: C. About 128 MHz
Frequency is proportional to field strength: 42.58 x 3 is about 128 MHz. 64 MHz applies to 1.5 T, 43 MHz to 1 T, and 255 MHz to 6 T.
Reference: Bushberg et al., The Essential Physics of Medical Imaging, 4th ed., 2021
Question 6 · Practical Experience
A pressure gauge used to calibrate NIBP monitors is itself being calibrated. Which reference arrangement is acceptable?
- AAnother working gauge of the same make and age
- BThe device under test used as its own reference
- CA reference standard traceable to a national standard with a test uncertainty ratio of at least 4 to 1
- DA reference accurate to the same level as the gauge under test
Show answer and explanation
Correct answer: C. A reference standard traceable to a national standard with a test uncertainty ratio of at least 4 to 1
Calibration needs a traceable reference clearly more accurate than the unit under test, usually at least four times. A similar gauge, or the device itself, or an equal-accuracy reference cannot show the error reliably.
Reference: ANSI/NCSL Z540.3-2006, Requirements for calibration of measuring and test equipment
Question 7 · Biomedical Equipment Technology
A ventilator in volume control shows a high peak pressure alarm while the plateau pressure remains normal. Which cause fits?
- AReduced lung compliance, for example from pulmonary oedema
- BRaised airway resistance, such as a kinked tube
- CA leak in the breathing circuit
- DA disconnected patient circuit
Show answer and explanation
Correct answer: B. Raised airway resistance, such as a kinked tube
When the peak rises but the plateau does not, the problem is in the airways or tube, so resistance has risen. Poor compliance raises both pressures. A leak or a disconnection lowers the pressure and gives a low-pressure alarm.
Reference: Bronzino, The Biomedical Engineering Handbook, 4th ed. (2015)
Question 8 · Advanced Electronics and Electricity
A non-inverting amplifier has a 90 kilohm feedback resistor and a 10 kilohm resistor to ground. What is its gain?
- A9
- B-9
- C10
- D100
Show answer and explanation
Correct answer: C. 10
The gain of a non-inverting amplifier is 1 + Rf/Rg = 1 + 90/10 = 10. The value 9 forgets the 1, and -9 is the gain of an inverting stage. The value 100 does not follow from the formula.
Reference: Sedra and Smith, Microelectronic Circuits, 7th ed. (2014)
Question 9 · Anatomy and Physiology
An ECG simulator is set to a rhythm with no visible P waves and an irregularly irregular ventricular rate. Which rhythm is this?
- ASinus arrhythmia
- BThird-degree heart block
- CAtrial flutter with a fixed block
- DAtrial fibrillation
Show answer and explanation
Correct answer: D. Atrial fibrillation
Absent P waves with a chaotic ventricular rate are the pattern of atrial fibrillation. Sinus arrhythmia keeps P waves. Complete heart block has a regular slow ventricular rhythm, and flutter with a fixed block gives a regular rate with sawtooth waves.
Reference: Webster, Medical Instrumentation: Application and Design, 4th ed. (2010)
Question 10 · Chemistry
In the Spaulding classification, which item category requires sterilisation?
- ACritical items that enter sterile tissue, such as surgical instruments
- BSemi-critical items that touch mucous membranes, such as a bronchoscope
- CNon-critical items that touch intact skin, such as a blood pressure cuff
- DEnvironmental surfaces such as bed rails
Show answer and explanation
Correct answer: A. Critical items that enter sterile tissue, such as surgical instruments
Items that penetrate sterile tissue or the vascular system carry the highest infection risk and must be sterilised. Semi-critical items need at least high-level disinfection. Non-critical items need low-level cleaning or disinfection, and surfaces are cleaned.
Reference: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008)
Question 11 · Professionalism
A ventilator that delivered an excess pressure has injured a patient. What should the technician do first?
- ARepair and recalibrate it before the review so it returns quickly
- BQuarantine it with settings and logs intact, then report the event to the manager
- CClear the event log to remove the alarm history
- DReturn it to the vendor without telling anyone
Show answer and explanation
Correct answer: B. Quarantine it with settings and logs intact, then report the event to the manager
The device is evidence and must be quarantined with settings and logs intact so that the cause can be found, and the event must be reported. Repairing it, clearing the log or sending it away destroys evidence and delays learning.
Reference: Dyro, Clinical Engineering Handbook (2004)
Question 12 · Patient Safety, Electrical Safety and Hazards
Which action best reduces fire risk during head and neck surgery that uses electrosurgery?
- ARaise the oxygen concentration around the surgical site
- BAllow alcohol prep solution to pool under the drapes
- CLet skin prep dry and keep oxygen as low as is safe
- DRun the unit at its highest power for most cutting
Show answer and explanation
Correct answer: C. Let skin prep dry and keep oxygen as low as is safe
A fire needs an ignition source, fuel and an oxidiser. Wet alcohol is fuel and extra oxygen makes the fire burn faster, so both are limited. Pooled solution and unnecessarily high power increase the hazard.
Reference: ASA Practice Advisory for the Prevention and Management of Operating Room Fires (2013)
Biomedical Engineering & Medical Devices reference books and guidelines
Every question in this bank names the textbook or guideline its answer rests on. These are the sources our Biomedical Engineering & Medical Devices questions cite most often - the reading list to revise from alongside the bank.
Textbooks
8- 1
Webster (ed.), Medical Instrumentation: Application and Design
4th edition2010 - 2
Horowitz and Hill, The Art of Electronics
3rd edition2015 - 3
Guyton and Hall, Textbook of Medical Physiology
14th edition2020 - 4
Bushberg et al., The Essential Physics of Medical Imaging
4th edition2021 - 5
Khandpur, Handbook of Biomedical Instrumentation
3rd edition2014 - 6
Dyro (ed.), Clinical Engineering Handbook
2004 - 7
Dyro (ed.), Clinical Engineering Handbook,, ventilator chapter
2004 - 8
Bland and Altman, Lancet;1:307-310
1986
Guidelines and standards
18- 1
IEC 60601-2-4:, Particular requirements for cardiac defibrillators
2010 - 2
ISO 81060-2:, Non-invasive sphygmomanometers, clinical validation
2018 - 3
AAMI ST79:, Comprehensive guide to steam sterilization
2017 - 4
CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities
2008 - 5
NEMA, DICOM PS3, current edition
- 6
HL7 International, FHIR Release 4 (v4.0.1)
2019 - 7
NEMA, DICOM PS3.4, current edition
- 8
IEC 80001-1:, Application of risk management for IT networks incorporating medical devices
2021 - 9
IEC 62353:, Recurrent test and test after repair of medical electrical equipment
2014 - 10
IEC 62353:; IEC 60601-2-4
2014 - 11
ANSI/NCSL Z540.3-, Requirements for calibration of measuring and test equipment
2006 - 12
IEC 60601-1:+AMD1:+AMD2
2020 - 13
IEC 60601-2-2:, High frequency surgical equipment
2017 - 14
IEC 62353
2014 - 15
ASTM F2503-20, Marking medical devices for safety in the MR environment
- 16
IEC 60601-1-8:+A1:+A2:; Joint Commission NPSG.06.01.01
2020 - 17
SFDA, Medical Devices Interim Regulation (MDS-IR), current edition
- 18
ISO 14971:, Medical devices, application of risk management
2019
Where your regulator publishes a reading list, use the editions it names. Exam questions are not taken from any one book.
Last reviewed September 2026
About these questions: every question is original, researched by Neelim's exam research team, written to the published blueprint, with a rationale and a reference. No recalled or leaked exam content is used. How we write our questions.
Before you book your exam
Frequently asked questions
Are these real Biomedical Engineering & Medical Devices exam questions?