Massive Air Embolism and Arterial Line Emergencies
You are the primary perfusionist on a mitral valve repair in an Abu Dhabi cardiac centre, on full bypass at 32 degrees with a roller arterial pump at 4.6 L/min. The surgeon lifts the heart, venous return falls suddenly, and the reservoir empties before you react; the level sensor had been switched off during priming and not reactivated. You see a column of air moving down the arterial line towards the aortic cannula. What do you do?
What the panel is testing: Whether you can run the massive air embolism sequence in the right order under pressure, lead the team through it, and recognise the safety failure behind it.
Show the answer outline
Knowledge: Air pumped into the arterial line goes straight to the aorta and the cerebral and coronary circulations, causing stroke, myocardial ischaemia and death. With a roller pump, an empty reservoir lets the pump drive air forward at full flow, which is why a level sensor linked to the pump is a basic safety requirement. The classic response, described by Mills and Ochsner, is to stop the pump, clamp the lines, de-air the circuit and the patient, and use retrograde perfusion through the superior vena cava to flush air out of the cerebral vessels. Hypothermia lowers cerebral metabolic demand, and hyperbaric oxygen may help later if available.
Management: I would stop the arterial pump at once, call 'air in the arterial line, stop', and clamp the arterial and venous lines. The surgeon places the patient in steep head-down position, and the aortic cannula is removed or opened so air can be vented from the aorta and arterial line. I would de-air and refill the circuit, re-prime the arterial line and filter, and restore the level. Retrograde perfusion of the superior vena cava with cold oxygenated blood is then run, as the surgeon directs, until no more air comes from the aortic root. Antegrade bypass then resumes with deeper cooling and adequate perfusion pressure,. After the case the patient needs ICU neurological assessment and possible hyperbaric referral.
Ethics and communication: I would record exact times, flows, temperatures and actions, and keep the circuit for inspection. The disabled level sensor is a serious incident that I report through the facility system and review honestly. The family should receive open disclosure from the surgeon with perfusion input, in line with Federal Decree-Law No. 4 of 2016 on Medical Liability. Pre-bypass checklists confirming that safety devices are active should be reinforced.
Likely panel follow-up questions
- Why does a roller pump make this more dangerous than a centrifugal pump?
- How would you perform retrograde superior vena cava perfusion with your circuit?
- What safety devices should be active before you go on bypass?
- How would the anaesthetist assess the patient for neurological injury afterwards?
Pitfalls in this case
- Pausing to diagnose instead of stopping the pump
- Restarting flow before the arterial line and aorta are de-aired
- Not communicating clearly to the surgeon and anaesthetist
- Failing to report the disabled level sensor
Reference: Mills NL, Ochsner JL, Massive air embolism during cardiopulmonary bypass (J Thorac Cardiovasc Surg 1980); Cardiopulmonary Bypass and Mechanical Support: Principles and Practice (5th ed, 2025)