Air Embolism and Extracorporeal Circuit Emergencies
You are the dialysis technologist on the afternoon shift in an Abu Dhabi outpatient unit. A 61-year-old man is 50 minutes into haemodialysis through a forearm fistula at a blood flow of 350 mL/min. The venous air detector has alarmed twice and a colleague reset it both times. You find the priming saline line unclamped and the bag empty. He is coughing, his chest feels tight, and his saturation has fallen from 97 to 86 percent with a pulse of 118. What do you do?
What the panel is testing: Whether you recognise venous air embolism at once, stop further air entry, and position and escalate the patient correctly.
Show the answer outline
Knowledge: An open saline line before the blood pump sits in the negative pressure segment, so an empty bag lets the pump draw air into the circuit. Alarms reset without finding the cause are the classic chain of events behind dialysis air embolism. Air reaching the venous return travels to the right heart and lungs and causes cough, chest tightness, breathlessness, hypoxaemia and arrhythmia. In a seated patient air can also rise into the cerebral veins, and a patent foramen ovale can let it cross to the arterial side and cause stroke-like signs. Other sources include a loose arterial needle or a catheter left open.
Management: I would clamp the venous line and stop the blood pump at once, and I would not return the circuit blood, because it may contain air. I would call the nurse in charge and the nephrologist and ask for the emergency trolley. He is placed on his left side with the head down, to hold air in the right atrium and away from the pulmonary outflow and the brain. I would give 100 percent oxygen by non-rebreather mask, check pulse, pressure, saturation and consciousness every few minutes, and start CPR if he loses a normal pulse. The physician decides on hospital transfer and hyperbaric oxygen. I would keep the lines intact for inspection and take the machine out of service.
Ethics and communication: I would document the times, alarm history and findings without altering the machine log. This is a serious incident for the facility reporting system, and the repeated alarm resets are a system issue for review, not a reason to blame one colleague. The patient and family should be told honestly what happened, in line with the open disclosure principle that underpins Federal Decree-Law No. 4 of 2016 on Medical Liability.
Likely panel follow-up questions
- Why is the arterial segment before the blood pump the dangerous place for air entry?
- What should you check every time the air detector alarms?
- He becomes confused with weakness of the right arm. What does that suggest?
- How would you prevent this happening again in your unit?
Pitfalls in this case
- Returning the circuit blood to the patient before clamping
- Resetting an air alarm without tracing the air
- Sitting the patient up because he feels breathless
- Not taking the machine out of service or not reporting the event
Reference: Daugirdas JT, Blake PG, Ing TS, Handbook of Dialysis (6th ed, 2025); Renal Association Clinical Practice Guideline on Haemodialysis (Ashby D et al, BMC Nephrol 2019)